MED-EL OPUS1 / OPUS2 / TEMPO+
with remote battery pack
List of required parts
1
MLxi or MicroMLxS
2
Remote Battery Pack
3
White cable
4
MicroLink CI S
5
Speech processor
Fitting with the user
1. Turn all equipment Off.
2. Connect the white cable to the MicroLink CI S and to the
Remote Battery Pack.
3. Plug in the FM receiver into the MicroLink CI S. When using
MicroMLxS, switch it to the double green dot position.
4. Turn the volume of the MicroLink CI S to the minimum.
5. Turn On the Remote Battery Pack and switch it into
MIX-Mode.
6. Turn On the MicroLink CI S and FM transmitter.
7. Gradually increase the gain of the MicroLink CI S to a
comfortable level.
8. Determine speech recognition ability in quiet with the
FM signal. Ensure that results are similar to implant
alone condition. MicroLink CI S gain may need further
adjustments.
9. Evaluate FM benefit as well as the appropriateness of FM
settings through speech-recognition assessment in the
clinic and standardized questionnaires completed by the
recipient, family members and teachers.
Listening check
The best way to listen to the FM signal is to plug the
MicroLink CI S with the blue patch cable into a loudspeaker.
Make sure there is a good battery inside the MicroLink CI S.
Then turn On the MicroLink CI S and FM transmitter.
By speaking into the transmitter’s microphone you should
hear your voice through the speaker.